Specifications for EU-M60
Back to main page Print
SpecificationsSystems

Power Supply Voltage 120V, 220-240V AC
Frequency 50 / 60Hz
Input Current 0.6A(120V), 0.3A (220-240V)
Voltage Fluctuation ±10%
Frequency Fluctuation ±?Hz
Size Dimensions( Main Unit ) 370mm(W) x 490mm(D) x 178mm(H)
Dimensions( Keyboard ) 390m(W) x 197mm(D) x 40mm(H)
Dimensions( Foot Switch ) 220mm(W) x 130mm(D) x 35mm(H).
Weight Main Unit 18.5kg
Keyboard 1.6kg
Foot Switch 1.5kg
Monitor Observation Display Mode B-mode
Scanning Display (Radial) Full circle, Bottom sector, Top sector, Scroll
Scanning Display (Linear) Dual display,Linear-priority display, Radial-priority display
Display Range 1, 2, 3, 4, 6, 9, 12cm range
Image Direction Normal / Inverse
Image Rotation (Radial) Rotatable ( 64 steps, Clockwise / Counterclockwise )
Image Rotation (Radial at DPR mode) Rotatable ( 512 steps, Clockwise / Counterclockwise )
Usable Frequencies C5, C7.5, C12, C20,
7.5MHz, 12MHz, 20MHz, 30MHz
Signal Processing Signal Processing Settings GAIN, STC and CONTRAST settings are retained in memory when power is tuned ON/OFF.
GAIN 20-step adjustable
Contrast 8-step adjustable
STC Partly amplifies/attenuates sections of the US image by using 7 segments in 9 steps.
IMAGE 4-step image adjustment of the ultrasonic image
Measurement Distance Measures distance defined by the + and x calipers.
Area/ Circumference Measures area/circumference enclosed by caliper tracing.
Volume Measures volume of area of interest in the DPR data obtained by herical scan, by tracing outline of the area on the each radial slice.
Display Hospital Name Displays the hospital name.
Calendar Timer Automatically displays date and time.
Patient Data Patient ID, name, D.O.B., age, sex.
Frequency Display Displays current frequency
Display Range Displays current range
Radial Direction Normal / Inverse
Herical (linear) Direction Normal / Inverse
Display Signal Processing Settings Displays GAIN, CONTRAST, STC
Stroke 15,20,30,40mm
Pitch 0.25,0.35,0.50,0.75, 1.00mm
Display Characters Alphanunmerics with Symbols.
Ancillary Equipment Photographic and Recording Units Video Printer( Color/Monochrome ), VTR,
Remote Controllers Foot Switch, Ultrasonic Endoscope Remote Switch
Others Keyboard Keyboard with built-in trackball.
Current Image Selectable The monitor display can be switched between endoscopic and ultrasonic images.
Preset Multiple image setting conditions can be stored.

Specifications and design are subject to change without any obligation on the part of the manufacturer.
Top
EU-M60
  Compatibility Charts
Software
Brochures