Specifications for CLV-180
Back to main page Print
Specifications

Automatic brightness control Method Servo-diaphragm
Automatic Exposure 17 steps
Air Feeding Pump Diaphragm
  Pressure switching  4 level available ( off, low, mid, high)
Water Feeding Method Air pressurisation or detachable water container
Indicators on Front Panel Emergency Lamp Reports absence, disconnection& use of lamp
NBI mode Reports preparation & use of NBI mode
Setting Memory  Settings (except for filter) are stored even when the light source is off 
 Illumination Examination Lamp Xenon short are lamp 300 W 
  Average lamp life Approx. 500 hrs on continuos use
  Ignition method Switching regulator
  Brightness adjustment Light-path diaphragm control
  Cooling Forced-air cooling
  Colour conversion Possible by using special purpose filter
  Emergency Lamp Halogen lamp( with mirror) 12 V 35 W
  Average emergency lamp life Approx. 500 hours
17 steps  

Classification as medical electrical equipment

 

Type of protection against electric shock  Class 1
  Degree  of protection against electric shock of applied part Type BF applied part. NOTE: The applied part as classification types that are not marked are Type BF applied parts
  Degree of protection against explosion Use prohibited in flammable environment 
Power Supply Voltage 100-120V, 220-240 V AC
  Voltage fluctuation Within +/- 10 %
  Frequency  50/60 Hz
  Frequency fluctuation  Within +/- 1Hz
  Consumption electric power 500 VA
  Fuse rating  8A, 250V
  Fuse size 5 mm x 20 mm
 Size Dimensions 381(W) x 162(H) x 536 (D) mm (maximum)
  Weight 15.4kg

Specifications and design are subject to change without any obligation on the part of the manufacturer 
Top